Automated external defibrillators

HeartStart FRx

Automated external defibrillator

Find similar products

The HeartStart FRx defibrillator includes advanced Life Guidance features to help guide the treatment of sudden cardiac arrest. With easy set-up, clear voice prompts, and rugged design, HeartStart FRx is designed for on-the-spot responders.

Specifications

Environmental - HeartStart FRx
Environmental - HeartStart FRx
Temperature
  • Operating/Standby:32° - 122° F (0° - 50° C)
EMI (Radiated/Immunity)
  • CISPR II Group I Class B, IEC 61000-4-3, and IEC 61000-4-8
Sealing
  • Waterjet proof IPX5 per IEC60529Dust protected IPX5 per IEC60529
Altitude
  • 0 to 15,000 feet
Aircraft
  • Device: RTCA/DO-160D;1997
Crush
  • 500 pounds
Vibration
  • Operating: meets MILSTD 810F Fig.514.5C-17, randomStandby: meets MILSTD 810F Fig.514.5C-18,swept sine.
Product specifications
Product specifications
Model Number
  • 861304
Therapy
  • Adult defibrillation peak current: 32A (150J nominal) into a 50 ohm load. Pediatric defibrillation (with optional Infant/Child Key installed): 19A (50J nominal) into a 50 ohm load.
Battery
  • Typically 4 years when stored and maintained according to directions provided in Owner’s Manual. Standby life (after insertion):
Waveform
  • Truncated Exponential Biphasic. Waveform parameters adjusted as a function of each patient’s impedance.
Protocol
  • Protocol Device follows preconfigured settings. Defibrillation and CPR protocol can be customized using HeartStart Event Review or HeartStart Configure software.
Quick Shock
  • Able to deliver a shock after the end of a CPR interval, typically in eight seconds.
Physical requirements
Physical requirements
Size
  • 2.4 x 7.1 x 8.9 inches (6 x 18 x 22 cm) H x D x W
Weight
  • With battery and pads case: 3.5 lbs (1.5 kg)Without battery or pads case: 2.6 lbs (1.2 kg)
  • *”Shortening the interval between the last compression and the shock even by a few seconds can improve shock success (defibrillation and ROSC)” – American Heart Association. American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122 (suppl 3): S706-S719.