Respironics Ventilator

Respironics V60

Ventilator

989805611761

Find similar products

The Respironics V60 ventilator is a microprocessor-controlled, bilevel positive airway pressure (BiPAP) ventilatory assist system that provides non-invasive positive pressure ventilation (NPPV) and invasive ventilatory support for spontaneously breathing adult and pediatric patients.

Features
Auto-Trak
Advanced NIV with our patented Auto-Trak technology

Advanced NIV with our patented Auto-Trak technology

Auto-Trak technology is designed to address the specific challenges of noninvasive ventilation. By providing auto-adaptive leak compensation, inspiratory triggering, and expiratory cycling, Auto-Trak delivers optimal synchrony in the face of dynamic leak and changing patient demand. Hospital modes and options include: AVAPS, PCV, CPAP with C-Flex, PPV and Auto-Trak Plus.
Automatic Mask Calibration
Save time with automatic mask calibration

Save time with automatic mask calibration

We support your NIV efforts by increasing the speed and ease of treatment initiation. V60’s pre-defined settings for Philips Respironics hospital masks help you save time. You select the setting, and the V60 automatically calibrates flow characteristics for better monitoring and therapy.
Flexible
Flexible, easy, upgradeable

Flexible, easy, upgradeable

Designed to include pediatric use and equipped with several modes, the V60 allows you to meet the specific needs of your patients. The high-resolution touchscreen makes operation easy and facilitates waveform interpretation. The internal 6-hour battery enables intra-hospital transport. And, Respi-Link* allows you to upgrade via the internet. *Not available in all markets

Specifications

General
General
Dimensions
  • 33.7 cm H x 39.4 cm W x 42.9 cm D (13.3"H x 15.5" W x 16.5" D)
Oxygen inlet pressure range
  • 276 to 600 kPa (40 to 87 psig)
Weight without optional battery
  • 10.0 kg (22 lb)
Weight with optional battery
  • 10.9 kg (24 lb)
Electrical - IntelliBridge EC40/80
Electrical - IntelliBridge EC40/80
AC power
  • 300 V·A
AC voltage
  • 100 to 240 VAC
Battery operating time
  • 6 hours in normal conditions
AC frequency
  • 50 to 60 Hz
Alarm adjustable range
Alarm adjustable range
Lo Vr (low tidal volume alarm)
  • OFF - 1,500 ml
Hi Vr (high tidal volume alarm)
  • 200 - 3,500 ml
LIP T (low inspiratory pressure delay time)
  • 5 - 60 s
HIP (high inspiratory pressure alarm)
  • 5 - 50 cm H₂O
LIP (low inspiratory pressure alarm)
  • OFF, 1 - 40 cm H₂O
Lo VE (low minute ventilation alarm)
  • OFF, 0.1 - 991 /min
Lo Rate (low respiratory rate alarm)
  • 1 - 89 beats per minute
Hi Rate (hi respiratory rate alarm)
  • 5 - 90 beats per minute
Waveform window
Waveform window
Pressure waveform
  • 0 - 50 cm H₂O
Flow waveform
  • 240-2401 /min BTPS
Volume waveform
  • 0 - 3,500 ml BTPS
Patient Data
Patient Data
Breath phase/trigger indicator
  • Spontaneous, Timed, Exhale
PIP
  • 0 to 50 cm H₂O
Tidal volume
  • 0 to 3000 ml BTPS
Patient/total leak
  • 0 to 200 l/min BTPS
Minute volume
  • 0 to 99.0 l/min BTPS
Respiratory Rate
  • 0 to 90 beats per minute
Ti/Ttot
  • 0 to 91 %
Patient Trigger
  • 0 to 100 %
Patient types
Patient types
Suitability
  • Pediatric (≥ 20 kg) and Adult
Settings
Settings
IPAP
  • 4 to 40 cm H₂O
Min P (AVAPS minimum IPAP)
  • 5 to 30 cm H₂O
Max E
  • 0 to 100 cm H₂O/l
Max P (PPV maximum pressure limit)
  • 5 to 40 cm H₂O
C-Flex
  • Off, 1 to 3
Oxygen percentage
  • 21 to 100 %
Respiratory rate
  • 4 to 60 beats per minute
AVAPS target tidal volume
  • 200 to 2000 ml BTPS
Max V (PPV maximum volume limit)
  • 200 to 3500 ml
Inspiratory time
  • 0.30 to 3.00 s
Max R
  • 0 to 50 cm H₂Ol/s
CPAP
  • 4 to 25 cm H₂O
Ramp time
  • Off, 5 to 45 minute(s)
EPAP
  • 4 to 25 cm H₂O
Rise time
  • 1 to 5
Triggering and cycling
  • Auto-adaptive (Auto-Trak) or Auto-Trak Plus (optional). Trigger: normal, +1 to +7. Cycle: -2, -1, normal, +1 to +6
PPV%
  • 0 to 100 %
Max P (AVAPS maximum IPAP)
  • 6 to 40 cm H₂O