SureSigns VS4
Vital signs monitor
This product is no longer available
This product is no longer available

SureSigns VS4

Vital signs monitor
This product is no longer available
Expand vital signs monitoring with Philips SureSigns VS4. Respond to changing patient conditions by choosing between frequent vitals and spot check mode. Prompt urgent action with QuickAlerts and simplify workflow with QuickCapture and QuickCheck tools.

Specifications
  • SureSigns VS4
    Dimensions
    26 (10.2) x 22 (8.6) x 14.5 (5.7) Cm/in.
    Weight
    3.1 (6.9) – exluding optional recorder Kg/lb
    Screen
    8.4 inch SVGA TFT-AM LCD display, 800 active pixels/line; resolution: 600 active lines/frame; viewing angle ±60 degrees
    Stepwise deflation pressure (Adult)
    Diastolic: 10-245 (1.3-32.7); Systolic: 30-270 (4.0-36.0); MAP: 20-255 (2.7-34.0) mmHg/kPa
    Stepwise deflation pressure (Pediatric)
    Systolic: 30-180 (4.0-24.0); Diastolic: 10-150 (1.3-20.0); MAP: 20-160 (2.7-21.3) mmHg/kPa
    Stepwise deflation pressure (Neonatal)
    Systolic: 30-130 (4.0-17.0); Diastolic: 10-100 (1.3-13.3); MAP: 20-120 (2.7-16.0) mmHg/kPa
    NBP measurement accuracy
    Maximum standard deviation: 8 mmHg; Maximum mean error: ±5mmHg
    Subsequent cuff inflation (in NBP interval mode only)
    The subsequent inflation pressure is determined automatically, depending on the previous measurement and patient type
    SpO2 measurement range
    0-100%; pulse rate measurement range 30-300 pbm. Accuracy depends on sensor
    SpO2
    Meets EN ISO 9919 standard
    Welch Allyn Predictive Temperature
    Probe Sites: Oral, rectal, axillary; Range 26.7-43.3°C (80-110°F); Accuracy ±0.1°C (±0.2°F) in monitoring mode
    Covidien Tympanic Temperature
    • Range: 33°C-42°C (91.4°F-107.6°F)
    • Accuracy (overall range): ±0.2°C (±0.4°F)
    • Measurement response time: ≤2
    Covidien Tympanic Temperature
    Equivalency modes: ear (no adjustment), oral (adjusted from ear), core (adjusted from ear), rectal (adjusted from ear)
    Battery
    Lithium ion: 10.8-11.1V (with a “202” form factor)
    Data Output
    HL7 format, via Ethernet port; serial data
    Patient type
    Adult, pediatric, neonatal