Pinnacle³ SmartEnterprise centralises your Pinnacle³ radiation therapy planning system and gives you exceptional flexibility. SmartEnterprise enhances accessibility, maintenance, and management of the system and makes remote access easy.
OEM model | Oracle* X6-2
|
CPU | One Intel® Xeon® E5-2643 v46-core 3.4 GHz processor
|
RAM | 64GB DDR4
|
Clustering software | Oracle Solaris Cluster (Sun Cluster)
|
Operating system | Oracle Solaris 11.3
|
Intel turbo boost and hyperthreading | Yes
|
Virtual CPUs for Pinnacle³ multi-threading | One (1) 8-core CPU - 16 Threads
|
HBA | Sun Storage Dual 16 Gb Fibre ChannelPCIe Universal HBA, Emulex
|
HBA ports and supported speeds | 2 ports per server node4 GB/s, 8 GB/s, 16GB/sMinimum recommendation of 800 IOPS with a targeted 10k write speed
|
Hard drives | 2x 600 GB 10K rpm 2.5-inch SAS-3 HDD
|
Ethernet ports | 4x 10,000/1,000/100 Mbps Base-T Ethernet ports
|
Required Ethernet switch ports (customer provided) | 1 or 2 for network connections (2 if using recommended redundant network configuration); 1 for ILOM port
|
Size | 1 UHeight: 42.6mm (1.7in)Width: 436.5mm (17.2in)Depth: 737.0mm (29.0in)Weight: 18.0kg (40.0lb)
|
Rack mounting | Slide rail kit and cable management arm included
|
Peak heat load | 897 BTU/hour
|
Peak power | 263
W
|
Operating environment | 5°C to 31°C (41°F to 95°F); 10%–90%, non-condensingUp to 3,000 m, maximum ambient temperature is derated by 1°C per 300 m above 900 m
|
Acoustic noise | 7.0 Bels A-weighted operating, 7.0 Bels A-weighted idling; 63.1 dBA operating, 60.5 dBA idling; systems must be acoustically isolated from staff work areas.
|
Power supply | Dual-redundant Sun 760W AC HE Gold Power Supplies100-240 VAC, 50 or 60 Hz; IEC 320-C13 power connector
|
Current draw | 2.4 A @ 110 VAC1.2 A @ 220 VAC
|
Power cords | Philips will supply power cords to match the local power outletPhilips will supply power cords to match the local power outlet, or jumper cables to fit IEC 320-C14 rack PDU power sockets.or jumper cables to fit IEC 320-C14 rack PDU power sockets.
|
Optical drive | DVD+/-RW SATA-based drive (for use by Philips technical support only)
|
OEM model | Oracle X6-2
|
Current draw | 4.0 A @ 110 VAC2.0 A @ 220 VAC
|
CPU | Dual Intel® Xeon® E5-2699 v422-core 2.2 GHz processor
|
RAM | 64GB DDR4: upgradable to 384
GB
|
Clustering software | N/A
|
Operating system | Oracle Solaris 11.3
|
Intel turbo boost and hyperthreading | Yes
|
Virtual CPUs for Pinnacle³ multi-threading | Two (2) 18-core CPUs - 72 Threads
|
HBA | N/A
|
HBA ports and supported speeds | N/A
|
Hard drives | 5x 1.2TB 10k RPM SAS-3 (RAID 5)
|
Ethernet ports | 4x 10,000/1,000/100 Mbps Base-T Ethernet ports
|
Required Ethernet switch ports (customer provided) | 1 or 2 for network connections (2 if using recommended redundant network configuration); 1 for ILOM port
|
Size | 1 UHeight: 42.6mm (1.7in)Width: 436.5mm (17.2in)Depth: 737.0mm (29.0in)Weight: 18.0kg (40.0lb)
|
Rack mounting | Slide rail kit and cable management arm included
|
Peak heat load | 1,497 BTU/hour
|
Peak power | 550
W
|
Operating environment | 5°C to 31°C (41°F to 95°F); 10%–90%, non-condensingUp to 3,000 m, maximum ambient temperature is deratedby 1°C per 300 m above 900 m
|
Acoustic noise | 7.0 Bels A-weighted operating, 7.0 Bels A-weighted idling; 63.1 dBA operating, 60.5 dBA idling; systems must be acoustically isolated from staff work areas.
|
Power supply | Dual-redundant Sun 760W AC HE Gold Power Supplies100-240 VAC, 50 or 60 Hz; IEC 320-C13 power connector
|
Power cords | Philips will supply power cords to match the local power outlet, or jumper cables to fit IEC 320-C14 rack PDU power sockets.
|
Optical drive | DVD+/-RW SATA-based drive (for use by Philips technical support only)
|